3 System Preparation
Operator’s Manual
3 - 3
•
DO NOT connect this system to outlets with the same circuit breakers and
fuses that control the current to devices such as life-support systems. If this
system malfunctions and generates overcurrent, or when there is an
instantaneous current at power ON, the circuit breakers and fuses of the
building’s supply circuit may be tripped.
3.3
Powering ON/OFF
CAUTION
To ensure safe and effective system operation, you must perform daily
maintenance and checks. If the system begins to function improperly,
immediately stop scanning. If the system continues to function improperly, fully
shut down the system and contact the Mindray Customer Service Department
or a sales representative. If you use the system in a persistent improperly
functioning state, you may harm the patient or damage the equipment.
3.3.1 Checking Before Powering ON
Check before the system is powered on:
•
The temperature, relative humidity and atmospheric pressure meet the requirements of the
operating conditions. For details, see “2.4.2 Environmental Conditions”.
•
There is no condensation.
•
There is no distortion, damage or dirt on the system and peripheral devices.
If any dirt is found, cleaning shall be performed. For details, see “19 System Maintenance”.
•
There are no loose screws on the monitor, control panel.
•
There is no cable damage (e.g., power cord). Maintain secure connections to the system at all
times.
•
The probes and probe cables are free from damage or stains.
For details of probe cleaning and disinfection, see “17.1.5 Probes Cleaning and Disinfection/
Sterilization”.
•
No miscellaneous odds and ends are attached or affixed to the control panel.
•
Ensure that all connections are free from damage and remain clear of foreign object blockages.
There are no obstacles around the system and its air vent.
•
Probe cleaning and disinfection.
•
The entire scanning environment and field must be clean.
•
The locking mechanism of the casters can work normally.
3.3.2 Powering the System ON
Before getting the system started, check if the system is plugged in – verify that the system circuit
breaker is at “On”. After the power indicator on the control panel turns on, press the power button
to power the system on. The system image appears.
If access control has been set by the system administrator, you can access data in the system only
after logging onto the system. For details, see “4.1.9 Access Control”.
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...