7 3D/4D
Operator’s Manual
7 - 7
Move the probe across the body surface.
The arrow in the figure below indicates the movement of the probe.
You can move the probe in the opposite direction to the arrow.
7.3
Static 3D
Static 3D provides single frame image acquisition of 3D images. During the scanning; the probe
performs the scanning automatically.
7.3.1 Basic Procedures for Static 3D Imaging
Perform the following procedure:
1.
Select the proper probe and exam mode; make sure there is sufficient gel on the probe for
scanning.
2.
Obtain a 2D image, and optimize the image if necessary.
3.
Press
to enter Static 3D mode, and define the ROI and curved VOI.
Methods to adjust the ROI:
–
Under acquisition preparation status: Roll the trackball to change the ROI size, position
and curved VOI, press the right <Set> key to toggle among setting the ROI position, size
or curved VOI. Press the left <set> key to adjust VOI angle.
–
Tap [Flip VOI] to flip VOI. Tap [Reset VOI] to reset the angle of VOI to the original
status.
–
Enter touch screen mapping mode.
Draw a circle on the desired area. Move ROI over the circle to adjust the size and position
of ROI. Roll the trackball or press <Set> to adjust ROI.
1
ROI
2
Draw a circle here
3
ROI cover this area
3
1
2
澳
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...