7 3D/4D
Operator’s Manual
7 - 21
5.
Adjust the size and position of the sample box to contain the interested region within the
sample box.
However, the box should be small enough to contain only the heart, not the whole thorax.
In order to archive a good result, try to adjust the volume box and the sweep angle to be as
small as possible. The longer the acquisition time, the better the spatial resolution will be.
6.
Press <Update> to begin image acquisition.
–
Keep the probe still, the volume acquisition proceeds automatically and the acquired
images are displayed.
–
To avoid external interference to the image, make sure that there is minimal movement of
the mother.
–
In case of motion of the mother, the fetus or probe position change, cancel the current
acquisition and begin for a new acquisition.
After the calculation, the estimated heart rate is shown on the screen.
7.
Judge if the shown estimated heart rate is reasonable.
–
If the result is not reasonable, tap [REJECT HR] to perform the acquisition again.
–
Tap [ACCEPT HR] to accept the result and go to image review.
–
Tap [REJECT HR] to reject the HR but save the volume data.
Operations in review mode are the same as that of 4D cine review mode; please refer to the
details in 4D chapter.
After STIC acquisition, press <Set> to switch between cine auto play mode and manual mode.
7.8
iLive
iLive brings you a better imaging experience by adding a light rendering effect to the traditional
method. It supports the point lighting mode, parallel lighting mode as well as the torch lighting
mode, allowing human tissue texture to be revealed more clearly.
NOTE:
iLive is available under Smart 3D, Static 3D and 4D modes.
To Activate iLive
Perform the following procedure:
1.
Enter 3D/4D image viewing status, or double-click the saved 3D/4D cine file in the iStation or
Review screen.
2.
Tap [iLive] under the [Adv.] page on the touch screen to turn the function on, and adjust the
parameters.
Imaging using iLive
Perform the following procedure:
1.
Select the imaging mode:
–
Use the ordinary probe and press
to enter Smart 3D.
–
Or, use 4D probe and press
to enter 4D mode or Static 3D mode.
2.
Select the render mode to be iLive by touching the touch screen, and set the related parameters
(quality and angle, etc.).
3.
Press <Update> to begin acquisition.
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...