17 Probes and Biopsy
Operator’s Manual
17 - 55
NGB-035
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put the sterile sheath on the probe.
b.
Select a proper needle-guided bracket, and match the locating groove with the tab of the
transducer. Mount the bracket onto the transducer.
1
Clamp
2
The clamping nut of left/right clamp
3
The locating position block
4
Groove
5
V-shaped guiding block
6
The pressing block for separating
needle
7
Needle guide hole
8
Needle clamping cover of needle-
guided bracket
9
Needle adjusting nut of needle guided
bracket
10
Angle adjusting base
11
Angle mark
12
Angle locking nut
13
Angle adjusting block
14
Buckle
<1>
<4>
<3>
<2>
<5>
<6>
<7>
<8>
<9>
<10>
<11>
<12>
<13>
<14>