19 System Maintenance
Operator’s Manual
19 - 7
2.
The power indicator of
the monitor is light on,
but image is blank.
• The interval between
turning off and restarting
the system is too short –
wait a minimum time of 20
seconds.
• The monitor brightness or
contrast may be
improperly set.
• Turn off the system and wait
a minimum time of 1 minute,
and then restart the system.
• Adjust the monitor
brightness and contrast back
to factory defaults.
3.
The monitor displays
the characters and
menus but no images.
• The transmission power,
overall gain or TGC
controls are improperly
set.
• Check if a probe is
connected and/or fully
connected.
• The system is in the frozen
status.
• Adjust the transmission
power, gain or TGC control.
• Ensure proper probe
connection.
• Unfreeze the image.
4.
The image quality is
degraded
• The exam mode is not
correct.
• The settings of the image
post processing are not
correct.
• The image presets are
inappropriate.
• Select an appropriate exam
mode.
• Adjust the settings of the
image via post processing or
reset the post processing to
the default values.
• Reset the factory default
presets.
5.
The button is
responseless with the
system buzzing
There is too much dirt
blocking the button.
Check the control panel for the
button being blocked and press it
several times to release it.
Clean the button.
No.
Failure
Cause
Measure
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...