17 - 38
Operator’s Manual
17 Probes and Biopsy
2.
Remove the needle-guided bracket:
Hold the transducer in the left hand; unscrew the locking nut with the right hand to loosen the
retaining clamp. Raise the needle-guided bracket to separate the locating clamp/ bulge from
the locating grooves.
NGB-022
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
b.
Hold the probe by one hand, select the proper needle-guided bracket, and hold it with the
other hand. Match the groove of the bracket with the tab of the probe. Amount the bracket
onto the probe.
c.
Screw the pinch nut of the needle-guided bracket to confirm that the needle-guided
bracket is properly installed on the probe.
2.
Adjust the needle angle to the proper shift as required:
a.
Loosen the angle pinch nut.
1
Groove
2
Clamp
3
Needle type adjusting base
4
Needle type dial scale
5
Needle fixing nut
6
V-shaped cover
7
Angle adjusting base
8
Angle shift sign
9
Angle pinch nut
10
Angle block
11
Pinch nut
1
2
3
4
5
6
7 8 9 10 11
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...