17 - 54
Operator’s Manual
17 Probes and Biopsy
3.
Insert the biopsy needle:
a.
Adjust the dial scale to the required needle type shift.
b.
Pull the lock pin and close the V-shaped cover to fix the lock pin in the groove of the
needle type adjusting base, so as to install the needle into the guiding hole.
4.
Release the needle from the bracket:
a.
Pull the lock pin out until the V-shaped cover can be turned and opened up.
b.
Turn over the V-shaped cover to expose the needle. Remove the probe and bracket.
5.
Remove the needle-guided bracket:
a.
Unscrew the pinch nut to release the needle-guided bracket.
b.
Hold the probe and take out the bracket.