17 Probes and Biopsy
Operator’s Manual
17 - 67
NGB-054
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
b.
Hold the probe in one hand, select the correct needle-guided bracket and hold it with the
other hand. Match the inside of the bracket with the outside of the probe respectively.
Mount the bracket onto the probe. Rotate the locking nuts of the guided bracket at the
right and left sides to fix the bracket and the probe. Rotate the needle-shaped adjusting nut
to the ultimate position as shown in the figure.
1
Angle adjusting block
2
Angle locking nut
3
Right clamp of the bracket
4
Needle-shaped adjusting nut
5
Needle clamping cover
6
Left clamp of the bracket
7
Angle adjusting base
8
Locking nuts of the guided bracket at
the right and left sides
9
Torsion spring of the needle pressure cover 10
Biopsy needle pressure position
1
2
3
4
5
6
7
8
9
10
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...