Measurement, Comments and
Body Mark
14.1 Measurement
There are general measurements and application measurement. You can perform measurements on
a zoomed image, cine reviewing image, real-time image, or frozen image. For measurement details,
please refer to the
“Advanced Volume”.
WARNING
•
Be sure to measure areas of interest from the most optimal image plane to
avoid misdiagnosis from inaccurate measurement values.
•
To obtain accurate Doppler flow measurement values, make sure the
transmitting beam is not perpendicular to the flow, otherwise false readings
and potential misdiagnosis may result.
CAUTION
•
If an image is unfrozen or the mode is changed during a measurement, the
calipers and measurement data will be cleared from the screen, but the
measurement data will be stored in the report.
•
If the system is turned off or [End] is selected during a measurement, the
data not saved will be lost.
•
In Dual-B imaging mode, the measurement results of the merged image
may be inaccurate. Therefore, the results are provided for reference only,
not for confirming a diagnosis.
Measurement Accuracy
•
Error of 2D Images
Parameter
Value Range
Error
Distance
Full screen
Within ±3%
Area
Full screen
Within ±7%.
Circ
Full screen
Within ±10%
Angle
Full screen
Within ±3%.
Volume
Full screen
Within ±10%.
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...