3 System Preparation
Operator’s Manual
3 - 9
3.7
Connecting/Disconnecting a Probe
CAUTION
•
Press <Freeze> to freeze an image or turn off the power of the system
before connecting/disconnecting the probe. Otherwise, system or probe
failure may occur.
•
When connecting or disconnecting a probe, place it in a proper position, to
prevent the probe from falling off or becoming damaged.
•
Hang the probe cable to the hanger located under the control panel to
avoid excessively bending and damaging the cable.
•
Only use the probes provided by Mindray. Aftermarket probes may result in
damage or cause a fire.
3.7.1 Connecting a Probe
WARNING
The probes, cables and connectors should be in proper operating order and
free from surface defects, cracks and peeling. Otherwise, this may lead to
electrical shock.
Perform the following procedure:
1.
Turn the knob horizontally. Keep the cable end of the transducer upwards, and insert the
connector into the port of the system, and then press in fully.
2.
Turn the lock handle 90° clockwise to lock it securely.
3.
Place the probe properly to avoid being treaded or wrapping around other devices. DO NOT
allow the probe head to hang free.
澳
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...