17 Probes and Biopsy
Operator’s Manual
17 - 29
•
Confirm that the needle-guided bracket is free of damage, deformation, stripping, malfunction,
loose, or missing parts.
•
Confirm that the needle-guided bracket is securely mounted in the correct position.
•
Select the proper needle according to the specification above, and adjust the needle shift to the
same specification of the selected needle.
NGB-007 (Metal)
1
Support of needle-
guided bracket
Used for installing the needle-guided bracket on the probe.
2
Groove and tab of the
needle-guided bracket
Respectively matched with the tab and groove of the probe.
3
Angle adjusting base
There are 3 types of angles available to be adjusted.
4
Angle shift sign (40°,
50°, 60°)
Matched with the biopsy angle (40°, 50°, and 60°).
5
Angle pinch nut
Used for fixing the angle lock at a chosen angle.
6
Angle block
Used for determining the angle of the biopsy; different
specifications of blocks can be used.
7
Guiding block
Used for installing biopsy needle; there are five specifications
of guiding blocks for different biopsy needles.
8
Specification of guiding
block (14G)
Matched with the corresponding biopsy needle (14G).
<1>
<2>
<8>
<7>
<9>
<6>
<10>
<3> <4> <5>
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...