17 - 36
Operator’s Manual
17 Probes and Biopsy
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
b.
Select a suitable needle-guided bracket and match the groove to the tab of the probe.
Mount the bracket onto the probe.
c.
Screw the pinch nut of the needle-guided bracket to ensure that the needle-guided bracket
is properly installed on the probe.
2.
Adjust the needle angle to the proper shift as required:
a.
Loosen the angle pinch nut.
b.
Adjust the angle block to the desired level.
c.
Tighten the angle pinch nut.
3.
Install the guiding block:
a.
Select a suitable guiding block and push it into the groove above the angle block
b.
Screw the block's nut to secure the block.
c.
Insert a biopsy needle with the same specification as that of the guiding block into the
guiding block hole.
4.
Release the needle from the bracket:
a.
Loosen the guiding block's nut and slightly move the guiding block in the direction of the
needle’s tail.
1
Pinch nut of bracket
2
Angle pinch nut
3
Angle block
4
Clamp
5
Groove
6
Angle-adjusting base
7
Angle shift sign
8
Guiding block specification
9
Guiding block
10
Needle guide hole
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...