17 - 78
Operator’s Manual
17 Probes and Biopsy
The guideline turns green if it is able to reach the lesion mark.
7.
Perform biopsy by the guideline.
The distance indicating box shrinks by the time needle tip goes towards the lesion mark. When
the tip reaches the lesion mark, the box turns green and merges with the tip.
For detailed conversion procedure description, refer to chapter “17.3.5 Screen Display in the
Guiding”.
8.
Save the single-frame and multi-frame image if necessary.
17.3.2 Installation
WARNING
•
Do not bend the needle during installation.
•
Improper installation of the magnetic navigator leads to incorrect space
navigation.
NOTE:
•
Dispose needle, sensor support, and sterile cover by requirements of the manufacturer. Keep
the bracket sensor.
•
Clean and disinfect the sensor by requirements of the manufacturer and do not bend the sensor.
•
Refer to the manufacturer manual for magnetic navigation installation and other information.
•
Make sure the dimensions of sensor and eTRAX needle is basically the same, or magnetic
navigation may be incorrect.
•
eTRAX needle sensor is matched with eTRAX needle and trocar.
vTRAX
Specifications:
•
Biopsy needle
1
Needle
2
Needle bracket
3
Needle sensor
4
Sterile cover
1
2
3
4
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...