17 Probes and Biopsy
Operator’s Manual
17 - 63
b.
Remove the transducer and the needle-guided bracket to separate the biopsy needle.
c.
Rotate the clamping nuts of the guided bracket on the right and left side to release the
bracket and the transducer. Remove the bracket from the transducer.
NGB-042
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put the sterile sheath on the probe.
b.
Needle-guided bracket groove corresponds to the convex point of the probe, and install
the needle-guided bracket onto the probe.
1
Needle-guided bracket grooves
2
Guiding block fixing frame
3
Left/right clamp pinch nut
4
Needle guiding hole
5
Guiding block
6
Left clamp
7
Right clamp
8
Left/right clamp pinch nut
<7>
<8>
<1>
<2>
<3>
<4>
<5>
<6>