6 - 24
Operator’s Manual
6 Image Acquisition
CAUTION
Curved anatomical M image in the operator’s manual that it is provided for
reference, not for confirming a diagnosis. Generally it should be compared with
other device or make a diagnosis by non-ultrasonic methods.
NOTE:
Only phased probe supports Free Xros CM.
Perform the following procedure:
1.
In real-time 2D mode, adjust the probe and image to obtain the desired plane.
2.
Tap [TDI] on the touch screen or the user-defined TDI key to obtain the image.
3.
Tap [Free Xros CM] on the touch screen or the user-defined key to enter the mode.
4.
Use the trackball to define the start point of the sampling line on the 2D image.
The cursor displays as +, and can be moved within the 2D image only.
5.
Press <Set> to fix the start point, and the digital number “1” is marked beside the point.
6.
Define the next point using the trackball and <Set> key (tap [Undo] on the touch screen to
cancel the current point and activate the preview point). The system updates the time-motion
curve in real time, and each point is marked with a number in sequence.
7.
Repeat Step 6 to finish the sampling line. Double press <Set> to finish the editing.
8.
You can edit the curve if needed:
a.
After finishing the sampling line, tap [Edit] on the touch screen. The cursor becomes the
icon .
b.
Move the cursor over the curve, press <Set> to activate the spot.
c.
Move the cursor to change the shape of the curve.
d.
Double press <Set> to finish the editing.
NOTE:
Tap [Delete] on the touch screen to remove the curve, and tap [Edit] to re-edit the curve.
9.
Adjust the parameters to obtain the desired tissue of Free Xros CM image. Then, save the
image.
10. Press <B> to exit.
6.8.3 Anatomical M Mode Parameters
In anatomical M mode, adjustable parameters are similar with these in M mode.
6.9
PW/CW Mode
PW (Pulsed Wave Doppler) mode or CW (Continuous Wave Doppler) mode is used to provide
blood flow velocity and direction utilizing a real-time spectral display. The horizontal axis
represents time, while the vertical axis represents Doppler frequency shift.
PW mode provides a function to examine flow at one specific site for its velocity, direction and
features; while CW mode proves to be much more sensitive to high velocity flow display. Thus, a
combination of both modes will contribute to a much more accurate analysis.
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...