17 Probes and Biopsy
Operator’s Manual
17 - 59
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
b.
Select a proper needle-guided bracket, and match the tab with the groove of the
transducer. Mount the bracket onto the transducer.
c.
Rotate the clamping nuts of the guided bracket on the right and left side to fix the bracket
and the transducer. Rotate the needle-type adjusting nut to the ultimate position as shown
in the figure.
d.
Hold the transducer. Press the biopsy needle pressure position to separate it from needle
guided V-shaped block.
e.
Put the needle into the needle guided-bracket, and the needle leans to V-shaped block.