17 - 20
Operator’s Manual
17 Probes and Biopsy
Storage and Transportation
When all examinations for the day have been completed, confirm that the probe is in good
condition. After disinfecting the probe, confirm that the probe is in good condition and stored in a
suitable place.
•
To prevent the probe from being damaged, DO NOT store it in locations where it may be
exposed to:
–
Direct sunlight or X-rays
–
Sudden changes in temperature
–
Dust
–
Excessive vibration
–
Heat generators
•
When the probe is sent to MINDRAY Customer Service Department or sales representative for
repair, be sure to disinfect it and keep it in the carrying case to prevent infection.
Store and transport the probes in the following ambient conditions:
Probe
Ambient
temperature
Relative humidity
Atmospheric
pressure
C5-1U
-20°C - 55°C
30% - 95%RH (no condensation) 700 hPa - 1060 hPa
SC8-2U
-10°C - 60°C
20% - 95%RH (no condensation) 700 hPa - 1060 hPa
V11-3HU
-20°C - 55°C
20% - 95%RH (no condensation) 700 hPa - 1060 hPa
D8-4U
-10°C - 50°C
10% - 85%RH (no condensation) 700 hPa - 1060 hPa
DE10-3U
-10°C - 50°C
30% - 95%RH (no condensation) 700 hPa - 1060 hPa
L14-6WU
-20°C - 55°C
30% - 95%RH (no condensation) 700 hPa - 1060 hPa
L11-3U
-20°C - 55°C
20% - 95%RH (no condensation) 700 hPa - 1060 hPa
SP5-1U
-20°C - 55°C
20% - 95%RH (no condensation) 700 hPa - 1060 hPa
SC6-1U
-20°C - 55°C
20% - 95%RH (no condensation) 700 hPa - 1060 hPa
L14-3WU
-20°C - 55°C
20% - 95%RH (no condensation) 700 hPa - 1060 hPa
SC5-1U
-10°C - 60°C
20% - 95%RH (no condensation) 700 hPa - 1060 hPa
L20-5U
-20°C - 60°C
15% - 90%RH (no condensation) 700 hPa - 1060 hPa
L9-3U
-20°C - 55°C
20% - 95%RH (no condensation) 700 hPa - 1060 hPa
L14-5WU
-20°C - 55°C
20% - 95%RH (no condensation) 700 hPa - 1060 hPa
P10-4U
-20°C - 55°C
30% - 95%RH (no condensation) 700 hPa - 1060 hPa
C6-2GU
-20°C - 55°C
20% - 95%RH (no condensation) 700 hPa - 1060 hPa
C11-3U
-20°C - 55°C
30% - 95%RH (no condensation) 700 hPa - 1060 hPa
P7-3U
-20°C - 55°C
30% - 95%RH (no condensation) 700 hPa - 1060 hPa
L16-4HU
0°C - 60°C
30% - 95%RH (no condensation) 700 hPa - 1060 hPa
L16-4Hs
0°C - 60°C
30% - 95%RH (no condensation) 700 hPa - 1060 hPa
C4-1U
-20°C - 60°C
15% - 90%RH (no condensation) 700 hPa - 1060 hPa
DE10-3WU
-10°C - 50°C
10% - 90%RH (no condensation) 700 hPa - 1060 hPa
CW5s
-20°C - 55°C
30% - 95%RH (no condensation) 700 hPa - 1060 hPa
CW2s
-20°C - 55°C
30% - 95%RH (no condensation) 700 hPa - 1060 hPa
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...