17 Probes and Biopsy
Operator’s Manual
17 - 83
7.
Perform needle navigation guiding; refer to “17.3.1 Basic Procedures for Needle Navigation
Guiding” for details.
8.
Anticlockwise eTRAX needle interface to take out the needle from trocar after the eTRAX
needle access the target tissue.
9.
Select the proper Biopsy/melt needle according to the actual situation.
17.3.3 Accuracy Verification
WARNING
•
This manual describes simple accuracy verification only. Perform elaborate
accuracy verification procedure based on clinical scenario if necessary.
•
If the needle is bent, projective line of the needle and projective point of the
tip will not suit the actual situation.
•
Be sure to conduct accuracy verification before needle navigation. Perform
needle navigation when the measured result is in the acceptable range.
Accuracy error is measured by comparing actual biopsy depth and the measured depth in the
ultrasound image. Needle navigation guidance feature should be activated only when the error is
acceptable.
Perform the following procedure:
1.
Connect all accessories of magnetic navigator.
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...