17 Probes and Biopsy
Operator’s Manual
17 - 37
b.
Separate the residual part of the needle-guide bracket and the probe from the needle.
5.
Remove the needle-guided bracket:
a.
Screw the pinch nut to release the needle-guided bracket
b.
Separate the bracket and the probe.
NGB-021
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the sterile probe sheath.
b.
Loose the retaining clamp, align the needle-guided bracket with the transducer to locate
the locating clamp on the needle guide to the locating grooves on the transducer, and then
press the retaining clamp (front) to match the locating clamp and the locating bulge on the
front retaining clamp to the corresponding groove.
c.
Screw the retaining clamp to match with the transducer structure. When the clamp is
turned to the right position, the locking nut will lock the retaining clamp and the needle-
guided bracket is then mounted to the right position.
1
Locating bulge
2
Needle guide
3
Retaining clamp (front)
4
Locating clamp
5
Retaining clamp (back)
6
Probe
7
Locking nut
8
Locating groove
<1>
<3>
<4>
<2>
<5>
<6>
<7>
<8>
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...