1 Safety Precautions
Operator’s Manual
1 - 5
•
Do not use a USB memory device (e.g., a USB flash drive, removable hard
disk) which has unsafe data. Otherwise, system damage may result.
•
It is recommended to only use the video devices specified in this manual.
•
Do not use gel, disinfectant, probes, probe sheath or needle-guided
brackets that are not compatible with the system.
•
The applied contrast agency should be compliant with the relevant local
regulations.
•
Read the Acoustic Output Principle in the operation manual carefully before
operating this system on clinical examination.
•
The cover contains natural rubber that can cause allergic reactions in some
individuals.
•
Please use the ultrasound gel compliant with the relevant local regulations.
•
Normal operation may be affected by unstable mains power supply; it is
recommended that our product be powered from an uninterruptible power
supply.
•
DO NOT expose the system to excessive vibration through transportation.
Mechanical damage may result.
•
Always keep the system dry. Avoid transporting this system quickly from a
cold place to a warm place; otherwise condensation or water droplets may
form allowing a short circuit and possible electric shock.
NOTE:
•
DO NOT use the system in the vicinity of strong electromagnetic field (such as a transformer),
which may affect the performance of the system.
•
Do not use the system in the vicinity of high-frequency radiation source (e.g. cellular phones),
which may affect the performance of the system or even lead to failure.
•
When using or placing the system, keep the system horizontal to avoid imbalance.
•
To avoid damaging the system, do not use it in following environment:
–
Locations exposed to direct sunlight.
–
Locations subject to sudden changes in environmental temperature.
–
Dusty locations.
–
Locations subject to vibration.
–
Locations near heat generators.
–
Locations with high humidity.
•
Turn ON the system only after the power has been turned OFF for a while. If the system is
turned ON immediately after being turned OFF, the system may not be rebooted properly and
could malfunction.
•
Select “Freeze” to freeze an image or turn off the power of the system before connecting or
disconnecting a probe.
•
Remove the ultrasound gel from the face of the transducer when the examination is completed.
Water in the gel may enter the acoustic lens and adversely affect the performance and safety of
the transducer.
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...