6 - 46
Operator’s Manual
6 Image Acquisition
No. Device
Name
Port
Description
1
Auxiliary
output
power
supply
cable
Power supply port Plug the auxiliary output power supply cable to the port of
the magnetic navigator, and connect the power supply.
NOTE:
Please use the auxiliary output power supply cable
equipped with the Ultrasound System.
2
USB data
cable
USB port
Plug one end of the USB data cable to the USB port of the
magnetic navigator, and another end to the USB port of the
Ultrasound System.
3
Magnetic
generator
Transmitter port
Plug the magnetic generator to the Transmitter port. Place it
around the patient’s shoulder or two sides of the patient’s
abdomen. The magnetic generator should be frontal-faced
with the lesion. The frontal face of the magnetic generator
keeps 15-60 cm from the target within the scope 30 cm up/
down and towards probe.
NOTE:
Keep the magnetic generator away from the metal
devices or electromagnetic interference.
4-9 /
Positioning
(probe) sensor port
Motion (abdomen)
sensor port
1. Plug two sensors to positioning (probe) sensor port
, and motion (abdomen) sensor port
respectively.
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...