17 - 34
Operator’s Manual
17 Probes and Biopsy
Perform the following procedure:
1.
Install the needle-guided bracket:
a.
Put on the probe cover.
b.
Hold the probe by one hand, select proper needle-guided bracket, and hold it with the
other hand, and align the grooves of the needle-guided bracket with the tabs of the probe,
then push the needle-guided bracket forward, making the grooves of the needle-guided
bracket to match with the tabs of the probe. Set the needle-guided bracket at the desired
position, turn tightly the knob of fixing needle-guided bracket to fix the needle-guided
bracket.
c.
Check manually to confirm the needle-guided bracket is securely installed on the probe.
d.
Select a proper guiding block and push it into the groove above the support of needle-
guided bracket, then turn tightly the knob of fixing the guiding block to fix the guiding
block on the support of needle-guided bracket.
e.
Insert a biopsy needle with the same specification as that of the guiding block into the
hole of the guiding block.
2.
Remove the needle-guided bracket:
a.
Remove the guiding block slightly along the direction of the needle's tail, and separate the
residual part of the needle-guide bracket and the probe from the needle.
b.
Remove the support of needle-guided bracket from the probe.
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...