3.1 Pre-Operation Inspection
23
9
3
To ensure safe use of the unit, be sure to check the following inspection
items prior to performing measurements.
Basic
Measurement
Chapter 3
3.1 Pre-Operation Inspection
Items
Countermeasure
See:
Inspect the unit, probe, and fixture
(Are there any damaged parts?)
If there is damage:
Unit and fixture: Submit them for repairs.
Probe: Replace it with a new one.
Inspect the connection cord
(Is the covering cracked or is any
metal exposed?)
Do not use a damaged cord because doing so
may result in electric shock. (Replace the cord
with a new one.)
Check the power supply voltage
setting
(Does the setting of the voltage
selector on the rear of the unit
match the power supply voltage to
be used?)
Use of the unit outside the specified power sup-
ply voltage range may result in the unit being
damaged or an electrical fault. Set the voltage
selector in accordance with the power supply
voltage to be used.
Setting the Voltage
Selector:
2.2 "Checking the
Power Voltage" (p. 18)
When the power is turned on,
does the fan spin and do the
“3504-40, 3504-50, 3504-60” and
version number indications ap-
pear on the MAIN display area?
If the fan does not spin or the “3504-40, 3504-
50, 3504-60” and version number indications
are not displayed, the unit may be malfunction-
ing. Submit it for repairs.
Does the CV LED light up when
measurement is performed with
the measurement terminals
open while the probe and fixture
are connected to the unit?
(Range: AUTO)
If the CV LED does not light up, the unit, probe,
or fixture may be malfunctioning.
Unit and fixture: Submit them for repairs.
Probe: Replace it with a new one.
Connecting the Probe
and Fixture:
2.4 "Connecting the
Probes and Fixtures" (p.
20)
Are the measurement values indi-
cated on the unit normal when
measuring known samples such
as standard capacitors?
If the measurement values are abnormal,
check/perform the following.
• Are the measurement conditions set appro-
priately?
• Perform open circuit and short circuit com-
pensation again.
• Turn load compensation off.
• Set an appropriate trigger delay time or wait
time for trigger synchronous output function.
If the measurement values are still abnormal af-
ter you have checked/performed the above, the
unit, probe, or fixture may be malfunctioning.
Unit and fixture: Submit them for repairs.
Probe: Replace it with a new one.
ment Conditions" (p. 26)
4.1 "Open Circuit Com-
cuit Compensation" (p. 37)
4.2 "Load Compensation"
(p. 47)
6.2 "Trigger Delay Setting"
(p. 93)
6.5 "Trigger Synchronous
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......