8.9 Message Reference
211
8
Setting and Query of OFFSET Compensation Function
Syntax
Command
Query
Response
:CORRection:OFFSet
<ON/ OFF>
:CORRection:OFFSet?
<ON/ OFF>
ON
:Enables the offset compensation function.
OFF
:Disables the offset compensation function.
Explanation
Command Set the offset compensation function to ON or OFF.
Query
Return the offset compensation function as ON or OFF.
Example
Command
:CORRection:OFFSet ON
Enables the offset function.
Query
Response
:CORRection:OFFSet?
:CORRECTION:OFFSET ON
(when HEADER ON)
ON
(when HEADER OFF)
The offset compensation function is enabled.
Setting and Query of OFFSET Compensation Values
Syntax
Command
Query
Response
:CORRection:OFFSet:DATA
<
C
offset compensation value
>,
<
D offset compensation value
>
:CORRection:OFFSet:DATA?
<
C
offset compensation value
>,<
D offset compensation value
>
<
C
offset compensation value
> = -99.9E -3 to 99.9E -3
<
The second parameter offset compensation value
>
= -1.99999 to
1.99999 (NR2)
Explanation
Command Sets the offset compensation value.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
Query
The offset compensation value settings are returned in the order of C
offset compensation value (NR3 numerical value), and D offset com-
pensation value (NR2 numerical value).
Example
Command
:CORRection:OFFSet:DATA 1.00000E-12,0.0001
Sets the C offset compensation value to 1.00000E-12 and the D offset
compensation value to 0.0001.
Query
Response
:CORRection:OFFSet:DATA?
:CORRECTION:OFFSET:DATA 1.00000E-12,0.0001
(when HEADER ON)
1.00000E-12,0.0001
(when HEADER OFF)
The C offset compensation value is set to 1.00000E-12, and the D off-
set compensation value is set to 0.0001.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......