5.2 BIN Measurement Function (Model 3504-50, 3504-60 only)
77
9
5
Setting the Upper Limit and Lower Limit Values for the BIN
Setting Workflow
1
Normal measurement
mode
Switch to comparator measurement mode.
Switch to BIN measurement mode.
The range is automatically set to HOLD.
2
BIN measurement mode
Switch to the setting mode for the upper limit
and lower limit values.
3
Reference value setting for C
(Only for the
% setting)
Confirm C.
Set the reference value for C.
Confirm the setting value.
4
Upper limit and lower limit
value settings for BIN1 to
BIN14
Confirm BIN1.
Set the upper limit value for BIN1.
Confirm the upper limit value.
Set the lower limit value for BIN1.
Confirm the lower limit value.
Set BIN2 to BIN14 in the same manner.
5
Reference value setting for D
(Only for the
% setting)
Confirm D.
Set the reference value for D.
Confirm the reference value.
Move to the next setting.
6
Upper limit and lower limit
value settings for D
Confirm D.
Set the upper limit value for D.
Confirm the upper limit value.
Set the lower limit value for D.
Confirm the lower limit value.
Move to the next setting.
7
BIN measurement mode
Switch to BIN measurement mode.
(Twice)
• Numeric
keypad
• Arrow keys
• Numeric
keypad
• Arrow keys
• Numeric
keypad
• Arrow keys
• Numeric
keypad
• Arrow keys
• Numeric
keypad
• Arrow keys
• Numeric
keypad
• Arrow keys
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......