6.8 Load the Measurement Conditions (Panel Load Function)
111
9
6
3.
Use the numeric keypad or
and
to select the panel number to
load.
The number is entered at the bottom of the MAIN display area.
4.
Press
to load the measurement conditions.
When the load condition is set to ALL (measurement conditions and compensa-
tion values) or hArd (measurement conditions), the unit switches to the saved
measurement mode.
When the load condition is set to Corr (compensation values), the unit returns to
the measurement mode it was in prior to the menu items being displayed.
• Only saved numbers can be set. If a number that was not saved with the
numeric keypad is set, the set number flashes and then changes to the
nearest saved number after one second elapses.
• At the time of shipment, when the unit has been reset, or at any other time
when there are no measurement conditions saved, “– –” is displayed at the
bottom of the MAIN display area.
• Each time a panel number is changed, the SUB display area for the mea-
surement conditions of that panel number flashes.
• The panel is not loaded unless
is pressed.
• After panel load is carried out, the self calibration value can be acquired by
setting the CALIb signal to Lo level in the response times (time from TRIG
(LOW) to INDEX (High) based on the load conditions in the “MANUAL”self
calibration setting.
*Response time in“ALL” and “measurement conditions” will vary according to
the time setting for the measurement speed.
The values in the table above obtained when the measurement speed is at
SLOW. During “circuit compensation value”, the response time will not vary
according to measurement conditions.
Load Condition
wait time* (ms)
ALL
350
measurement conditions
compensation value
0.5
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......