6.5 Trigger Synchronous Output Function
104
This function enables the measurement signal to be output after measurement is
triggered and ensures that the signal is applied to the sample only during mea-
surement. Thus reducing the generation of heat in the sample and decreasing
electrode wear.
1.
In normal measurement mode, press
.
The upper part of the MAIN display area displays the menu contents and the
lower part displays the setting information.
(Refer to "Menu display organization" (p. 13) for menu order)
This setting cannot be changed in comparator mode and BIN mode.
2.
Use
or
to select the “SYnC” menu item.
3.
Press
to enable or disable the Trigger synchronous output
function.
It will toggle “on” and “oFF” each time
is pressed.
4.
Press .
The trigger synchronous output function setting is confirmed.
When the trigger synchronous output function is set, the “SYNC SOURCE LED”
lights up.
After confirmation, “OFFSEt” (Offset settings screen) is displayed at the top of
the MAIN display area.
The trigger synchronous output function setting is not confirmed unless
is pressed.
6.5 Trigger Synchronous Output Function
(Trigger synchronous output function setting screen)
Menu content
“on” or “oFF” flashes
(MAIN display area)
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......