6.2 Trigger Delay Setting
93
9
6
This sets the delay time between when the trigger is detected and measuring
begins. With the use of this function even if measurement is commenced imme-
diately after connecting to a sample a reliable measurement value can be
attained.
1.
In normal measurement mode, press
.
The upper part of the MAIN display area displays the menu contents and the
lower part displays the setting information.
(Refer to "Menu display organization" (p. 13) for menu order)
This setting cannot be changed in comparator mode and BIN mode.
2.
Use
or
to select the “dELAY” menu item.
3.
Press
to enable or disable the trigger delay.
It will toggle “on” and “oFF” each time
is pressed.
4.
Press
to enable or disable trigger delay.
When “oFF” is selected, “SYnC” (Trigger synchronous output function setting
screen) appears in the MAIN display area.
When “on” is selected, the leftmost LED which displays the delay time will flash
in the middle row of the SUB display area.
6.2 Trigger Delay Setting
(Trigger delay setting screen)
Menu content
“on” or “oFF” flashes
(MAIN display area)
(SUB display area / “on” setting)
The delay time is lit up.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......