8.9 Message Reference
217
8
Setting and Query of Event Status Enable Register 3 (ESER3)
Syntax
Command
Query
Response
:ESE3
<Numeric Value>
:ESE3?
<Numeric Value>
<Numeric Value>
= 0 to 255 (NR1)
Explanation
Command Sets the mask pattern of the ESER3 to a numeric value from 0 to 255.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
The initial value (when the power is turned on) is 0.
Query
Returns the ESER3 content as an NR1 numeric value.
Example
Command
Query
Response
:ESE3 64
Sets bit 6 of ESER3.
:ESE3?
:ESE3 64
(when HEADER ON)
64
(when HEADER OFF)
Bit 6 of ESER3 is set to 1.
Note
There is no BIN function for the Model 3504-40.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
DNG
OUT
BIN14
BIN13
BIN12
BIN11
BIN10
BIN9
Query of Event Status Register 0
Syntax
Query
Response
:ESR0?
<Numeric Value>
<Numeric Value>
= 0 to 255 (NR1)
Explanation
Query
Returns the event status register 0 (ESR0) setting content as NR1
numeric data from 0 to 255 and then clears that content.
No header is added to the response message.
Example
Query
Response
:ESR0?
4
Bit 2 of ESR0 is 1.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
REF
COF
LER
MOF
MUF
IDX
EOM
CEM
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......