6.12 Printing Function
119
9
6
4.
Set the switches numbered 1 to 8 of DIP SW1 to either ON or OFF in
accordance with the table below.
Press the
ON LINE
switch once to set a switch to ON and the
FEED
switch
once to set a switch to OFF.
You can confirm the input result that is printed each time a switch is pressed.
If a setting is configured incorrectly, repeat the procedure from Step 1
.
is the setting to use with this unit.
After you finish configuring the switch numbered 8, the following is printed again.
Continue? :Push 'On-line SW'
Write?
:Push 'Paper feed SW'
5.
Press the ON LINE switch again so that the printer enters the configura-
tion state and configure each of the settings for DIP SW2 and DIP SW3
as shown in the table below.
Software DIP SW1 Settings
Switch
No.
Function
ON
(Press ON LINE)
OFF
(Press FEED)
1
Input method setting
Parallel
Serial
2
Print speed
Fast
Slow
3
Auto loading
Enable
Disable
4
CR function
Carriage return
Return
5
Setting command
Enable
Disable
6
Print density (set to
100%)
OFF
7
ON
8
ON
Software DIP SW2 Settings
Switch
No.
Function
ON
(Press ON LINE)
OFF
(Press FEED)
1
Print mode
Normal print
(40 digits)
Reduced print
(80 digits)
2
User-defined
character backup
Enable
Disable
3
Character type
Normal characters
Special characters
4
Zero font
0
Ø
5
International
characters
ON
6
ON
7
ON
8
ON
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......