8.9 Message Reference
174
(3) Synchronization Commands
Setting of OPC of SESR after All Executed Operations End
Syntax
Command
*
OPC
Explanation
Sets the OPC (bit 0) of SESR (standard event status register) when processing
ends for sent commands prior to the *OPC command.
Example
A;B;
*
OPC;C
Sets OPC of SESR after processing ends for commands A and B.
Response of 1 of ASCII after All Executed Operations End
Syntax
Query
Response
*
OPC?
1
Explanation
Responds with 1 of ASCII when processing ends for sent commands prior to the
*OPC command.
Continuing Execution of Commands after Command Processing Ends
Syntax
Command
*
WAI
Example
A;B;
*
WAI;C
Executes
*
WAI
and then the C command after processing ends for commands A
and B.
Current Frequency:1 kHz when in internal trigger state
• When the *WAI command was not used
(Send)
:FREQuency 120;:MEASure?
In this case, it is not certain which frequency measurement value will be sent in
response to the :MEASure? query.
• When the
*
WAI command was used
(Send)
:FREQuency 120;
*
WAI;:MEASure?
In this case, the 120 Hz frequency measurement value is sent in response to
the :MEASure? query.
Note
Unique commands other than the ":MEASure?" query use sequential commands.
Therefore, the
*
WAI command is only effective for the ":MEASure?" query.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......