5.2 BIN Measurement Function (Model 3504-50, 3504-60 only)
80
9.
Press
to enter the D upper limit value setting mode.
10.
Enter the D upper limit value using the arrow keys or the numeric
keypad.
Settable Range: OFF, -199999 to 199999 (When setting to "OFF"( p. 87))
Move to the digit ........
and
Change the number ..
and
Setting a value lower than the lower limit value will not produce an error, but
accurate judgment will not be possible.
At time of shipment, it is set to ” - - - - - - ” (OFF).
11.
Press
to confirm the D upper limit value.
Enter the D lower limit value setting mode.
12.
Enter the D lower limit value using the arrow keys or the numeric
keypad.
At time of shipment, it is set to ” - - - - - - ” (OFF).
13.
Press
to confirm the D lower limit value.
(SUB display area)
Light up
Flash
Upper limit value
(SUB display area)
Flash
Lower limit value
Light up