8.9 Message Reference
182
Setting and Query of Upper Limit and Lower Limit Values of First Parame-
ter for BIN Function in Count Value Mode
(Only for Models 3504-50, 3504-60)
Syntax
Command
Query
Response
:BIN:FLIMit:COUNt
<BIN Number>,<Lower Limit Value>,<Upper
Limit Value>
:BIN:FLIMit:COUNt?
<BIN Number>
<BIN Number>,<Lower Limit Value>,<Upper Limit Value>
<BIN Number>
= 1 to 14 (NR1)
<Lower Limit Value>
= OFF/ Numeric Value from -199999 to 999999 (NR1)
<Upper Limit Value>
= OFF/ Numeric Value from -199999 to 999999 (NR1)
Explanation
Command CommandSets the upper limit and lower limit values of the first
parameter in count value mode of the specified BIN number.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric value can be handled.
Query
Returns the upper limit and lower limit value settings for the first
parameter in count value mode of the specified BIN number in order
of BIN number, lower limit value, and upper limit value.
Example
Command
:BIN:FLIMit:COUNt 1,100000,150000
Sets 100000 for the lower limit value and 150000 for the upper limit
value of the first parameter in count value mode of BIN1.
Query
Response
:BIN:FLIMit:COUNt? 1
:BIN:FLIMit:COUNT 1,100000,150000
(when HEADER
ON)
1,100000,150000
(when HEADER OFF)
100000 is set for the lower limit value and 150000 is set for the upper
limit value of the first parameter in count value mode of BIN1.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......