4.2 Load Compensation
52
4.
Use the numeric keypad or arrow keys to enter a reference value
for C and then press
.
(If you enter a number, each digit moves one place to the right.)
Settable Range: -199999 to 999999
(excluding 0)
Move to the digit.........
and
Change the number ...
and
5.
Use the numeric keypad or arrow keys to enter a reference value
for D and then press
.
(If you enter a number, each digit moves one place to the right.)
Settable Range: -199999 to 199999
Move to the digit.........
and
Change the number ...
and
The unit returns to the state of Step 2.
(SUB display area)
Reference value of C
• If you do not want to change the reference value of C, press
with-
out changing the number. The reference value input screen for D is dis-
played.
• Set count values for the reference values. The reference values at the time
of shipment are 100000 for C and 0 for D.
(SUB display area)
Reference value of D
• If there is no need to change the reference value of D, press
with-
out changing the number.
• If you want to exit the reference value input screen and return to load com-
pensation setting screen, press
.
• Recalculates the load compensation rate if the reference value is changed
while LOAD compensation is enabled.
• Reference values are separated by measurement frequencies into separate
data. The reference value is set based on the current measurement fre-
quency.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......