8.9 Message Reference
184
Setting and Query of Reference Value of First Parameter for BIN Function in
Deviation Percent Mode (Only for Models 3504-50, 3504-60)
Syntax
Command
Query
Response
:BIN:FLIMit:REFerence
<Reference Value>
:BIN:FLIMit:REFerence?
<Reference Value> = Numeric Value from -199999 to 999999
(excluding 0) (NR1)
Explanation
Command Sets the reference value of the first parameter in deviation percent
mode. A numeric value in NRf format is accepted but non significant
digits are rounded off so the numeric value can be handled.
Query
Returns the reference value set for the first parameter in deviation
percent mode
Example
Command
:BIN:FLIMit:REFerence 150000
Sets 150000 for the reference value of the first parameter in deviation
percent mode.
Query
Response
:BIN:FLIMit:REFerence?
:BIN:FLIMit:REFERENCE 150000
(when HEADER ON)
150000
(when HEADER OFF)
150000 is set for the reference value of the first parameter in devia-
tion percent mode.
Note
The upper limit and lower limit values for count value mode and the
upper limit and lower limit values for deviation percent mode of the
unit are stored separately.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......