9.1 Basic Specifications
256
Self Calibration
Decrease the drift of the measurement value
OFF, MANUAL, AUTO
In MANUAL setting, when there is a request to start from EXT I/O during
the TRG signal activation, self calibration value will be acquired and ad-
justed.
When there is a request to start from communication, self calibration will
be acquired and adjusted in the measurement after the request for self
calibration.
Average
1 to 256 (necessary)
Triggerdelay
0 to 9.999s (0.001s resolution)
Contact Check Function
(only for Model 3504-60)
Result output during contact check abnormality
EXT I/O output ...error signal output (same signal line as CV error
signal)
LED display........error shown in MAIN display
See (p. 14)
Low C Reject Function
Detect contact abnormality (detects OPEN state of 2-terminal connec-
tion).
Output as a contact error if the measurement value is lower than the judg-
ment standard.
Judgement standard ..Can be set to 0.001% to 10.000% (0.001% resolu-
tion) of the full case of the range.
Error output ................LED lights up, as well as error output from EXT I/O.
Measurement Level
Monitoring Function
Detect contact abnormalities (detect chattering)
Takes the first waveform effective value (the first half of a waveform dur-
ing FAST setting, otherwise the first waveform effective value) as a basis
for comparison, and outputs a contact error if the fluctuation of the follow-
ing waveform effective values exceeds the judgment standard.
Judgment standard ....0.01% to 100.00% (0.01% resolution) of the refer-
ence value
Error output ................
output in the main display area, as
well as error output from EXT I/O
Trigger Synchronous Output
Function
Applying a measurement signal only during measurement is possible
Key Lock Function
The setting and canceling of this function by pressing a key on the front
panel is possible
BIN Measurement (Only for
Model 3504-50, 3504-60)
The setting of 14 categories for C and upper limit and lower limit values for
D is possible (absolute value setting,
% setting)
Comparator
The setting of the upper limit and lower limit values of each of C and D is
possible (absolute value setting,
% setting)
Product Specifications
Threshold value for contact check abnormality
Range
1 to 8
Range
9, 10
Hc terminal connection resi
Hp terminal connection resistance
Approx.
500
Approx.
100
Lc terminal connection resi
Lp terminal connection resistance
Approx.
500
Approx.
100
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......