4.1 Open Circuit Compensation and Short Circuit Compensation
44
Compensation Error:
If a compensation error occurs, a warning beep will sound and the state
becomes as follows.
The measurement frequency at which the error occurred can be ascertained
from the underbar display.
See
"MAIN display area ERROR display" (p. 14)
What if there is an error?
• If
is pressed, the unit returns to normal measurement mode. (The
settings for short circuit compensation remain the same as last time.)
• Are the measurement terminals in a short circuit state? Short circuit the
measurement terminals and then perform compensation again.
• If there is a compensation error even when the measurement terminals are
short circuited, the unit, probe, or fixture may be malfunctioning. Use the
shielding process, submit the unit or fixture for repairs, or replace the probe
with a new one. (The probe cannot be repaired.)
(MAIN display area)
Light up
Compensation stops.
When performing compensation, make sure that there is no noise source
nearby. Noise may cause an error when performing compensation.
ex. Servo Motor, switching power source, high-voltage cable and etc.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......