5.2 BIN Measurement Function (Model 3504-50, 3504-60 only)
81
9
5
14.
Press .
The unit switches to the BIN measurement mode.
See
"Performing BIN Measurement" (p. 88)
(Light up)
• The upper limit and lower limit values for the count setting become display
count values that are independent of the measurement conditions. If the
measurement conditions differ, the absolute values that signify the count
values change.
For example, if the range is currently set at 200 nF this would mean that C
count value of 50000 = 50E-9 and D count value 100 = 0.00100.
• The measurement conditions for normal measurement mode are used for
the BIN measurement mode. Set the measurement conditions to use for
BIN measurement mode while the unit is in normal measurement mode.
• Check the following because judgment cannot be performed properly if the
upper limit and lower limit values are set incorrectly.
• Are the setting values within the display range of the measurement
range?
• Is the large/small relationship of the upper limit and lower limit values cor-
rect?
• After settings are complete,
can be used to toggle the judgment
range display to confirm the judgment range.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......