4.4 Self Calibration
59
4
Performing Self Calibration ______________________________________
1.
In normal measurement mode, press
.
The upper part of the MAIN display area displays the menu contents and the
lower part displays the setting information.
(Refer to "Menu display organization" (p. 13) for menu order)
This setting cannot be changed in comparator mode and BIN mode.
2.
Use
or
to select the “CALIb” menu item.
3.
Use
and
to select a setting item.
Pressing
or
causes the display to change in the following manner
。
4.
Press .
The Self calibration setting is confirmed.
After confirmation, "IF" (Communication conditions setting screen) is dis-
played at the top of the MAIN display area.
The Self calibration setting is not confirmed unless
is pressed.
5.
Press .
(Self calibration settings screen)
Menu content
Self calibration
(set in step 3)
(MAIN display area)
“
MAnu
”
When inputting the trigger from EXT I/O and when the
CALIB signal line is at LOW, the self calibration value
will be calculated and adjusted during measurement.
Or when communication command (:CALibration:ADJust)
is executed, self calibration value will be calculated and
adjusted at the point of the first measurement after a
request for start.
“
Auto
”
Self calibration value will be calculated and adjusted
during each measurement.
“
oFF
”
Adjust using the factory default self calibration value.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......