8.9 Message Reference
191
8
Setting and Query of SUB Display Indication during Comparator
Measurement
Syntax
Command
Query
Response
:COMParator:DISPlay
<Character>
:COMParator:DISPlay?
<Character> = C/ D/ CREFerence/ DREFerence/ OFF
C
: Sets the upper limit and lower limit values of C to be displayed
in the SUB display area.
D
: Sets the upper limit and lower limit values of D to be displayed
in the SUB display area.
CREFerence: Sets the reference value of C to be displayed in the
SUB display area.
DREFerence: Sets the reference value of D to be displayed in the
SUB display area.
OFF : Sets nothing to be displayed in the SUB display area.
Explanation
Command CommandSets the set upper limit and lower limit values or the refer-
ence value to be displayed in the SUB display area during comparator
measurement.
Query
Returns the indication setting of the SUB display area during compar-
ator measurement as characters.
Example
Command
:COMParator:DISPlay C
Sets the upper limit and lower limit values of C to be displayed during
comparator measurement.
Query
Response
:COMParator:DISPlay?
:COMPARATOR:DISPLAY C
(when HEADER ON)
C
(when HEADER OFF)
The upper limit and lower limit values of C are set to be displayed dur-
ing comparator measurement.
Note
If an attempt is made to set the indication setting to CREFerence or
DEFerence when the judgment mode is count value mode, an execu-
tion error is generated.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......