8.9 Message Reference
201
8
Setting and Query of Open Compensation Points
Syntax
Command
Query
Response
:CORRection:OPEN:POINt
<Numeric Value>
:CORRection:OPEN:POINt?
<Numeric Value>
= 1 to 255 (NR1)
Explanation
Command Set measurement conditions for open compensation value acquisition
during command for open compensation value (:CORRection:OPEN
ALL) or acquiring open compensation value using key operation.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
Unused bit (bit6, bit 7) values will be disregarded.
In 3504-40 and 3504-50, bit0 and bit3 values are ignored.
Query
Returns settings of measurement conditions for open compensation
values.
Example
Command
:CORRection:OPEN:POINt 36
Set to acquire open compensation values under the measurement
conditions in “Yes” below during command for open compensation
value (:CORRection:OPEN ALL) or acquiring open compensation
value using key operation.
Query
Response
:CORRection:OPEN:POINt?
:CORRECTION:OPEN:POINT 36
(when HEADER ON)
36
(when HEADER OFF)
Returns the open compensation points.
Note
When any setting of the open compensation points is changed, all
open compensation values prior to the change will be disabled.
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
Unused
Unused
1 kHz,
1 V
1 kHz,
0.5 V
1 kHz,
0.1 V
120 Hz,
1 V
120 Hz,
0.5 V
120 Hz,
0.1 V
Measurement frequency
120 Hz
100 kHz
Measurement signal level
100 mV
No
No
500 mV
No
No
1 V
Yes
Yes
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......