8.9 Message Reference
227
8
Note
• If panel load has not been performed, or if measurement conditions
have been changed after panel load has been performed, the panel
load number will return 0.
Explanation
Query • Normal
measurement
Returns the measurement status, measurement values, and panel
load number.
• Comparator measurement
Returns the measurement status, measurement values, compari-
son result of the comparator, and panel load number.
However, if not even one measurement has been performed since
the measurement mode was changed, 2 is returned for the com-
parison results of both C and D.
• BIN measurement
Returns the measurement status, the measurement values, BIN
measurement result, and panel load number.
However, if not even one measurement has been performed since
the measurement mode was changed, OUT OF BINS is returned
for the BIN measurement result.
Example
Query
Response
Query
Response
Query
Response
• During normal measurement, when the second parameter is D.
:MEASure?
0,CP 1.23456E-06,D 0.12345,0
(when HEADER ON)
0,1.23456E-06,0.12345,0
(when HEADER OFF)
• Comparator measurement, the second parameter is D, when panel
number 5 is loaded.
:LOAD 5
:MEASure?
0,0,CP 1.23456E-06,0,D 0.12345,-1,5
(when HEADER ON)
0,0,1.23456E-06,0,0.12345,-1,5
(when HEADER OFF)
Indicates that the judgment result for C is IN and the judgment
result for D is LO.
• During BIN measurement, when the second parameter is D.
:BIN ON
:MEASure?
0,1,CP 1.23456E-06,D 0.01234,0
(when HEADER ON)
0,1,1.23456E-06,0.01234,0
(when HEADER OFF)
Indicates that the measurement value is within the set BIN1 range.
Note
• The header of C returns CS when the equivalent circuit is a series-
equivalent circuit and CP when the equivalent circuit is a parallel-
equivalent circuit.
Query of Measurement Data
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......