4.1 Open Circuit Compensation and Short Circuit Compensation
38
• The measurement accuracy values defined in the specifications are for when
open circuit compensation and short circuit compensation are performed.
• Be sure to perform compensation again after replacing the probe or fixture.
You will be unable to obtain correct values if measurement is performed in
the compensation state prior to replacement.
• The open circuit compensation range of impedance is 1 k
or more. How-
ever, if the values are not sufficiently high compared to the impedance of the
sample, the measurement errors will be larger and measurement may
become no longer possible.
• The short circuit compensation range of impedance is less than 1 k
. How-
ever, if the values are not sufficiently low compared to the impedance of the
sample, the measurement errors will be larger and measurement may
become no longer possible.
• The compensation value is saved at varying values depending on the mea-
sured frequency and signal level settings.
If these settings are changed and a measurement value has not been taken
with the changed measurement conditions, open circuit compensation and
short circuit compensation will be turned OFF.
• If there is an abnormality in the compensation value, measurement value
error will increase. The measurement value may be displayed as a negative
number.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......