5.1 Comparator Function
66
10.
Press
to set the D upper limit value.
Enter the D lower limit value setting mode.
11.
In the same way input the D lower limit value using the arrow keys
or the numeric keypad.
At time of shipment, it is set to ” - - - - - - ” (OFF).
12.
Press
to set the D lower limit value.
13.
Press .
The unit switches to the comparator measurement mode.
The comparator measurement result is displayed in the MAIN display area.
See
"Comparator Measurement Results" (p. 73)
Flash
(SUB display area)
Lower limit value
(Light up)
• The upper and lower limit values set during count setting become displayed
count values independent of measurement conditions. If the range settings
are changed, the absolute values which define the count values are
changed.
For example, if the range is currently set at 200 nF this would mean that C
count value of 50000 = 50E-9, D count value 100 = 0.00100.
• Set the comparator measurement mode measurement conditions while in
normal measurement mode.
See
3.3 "Setting the Measurement Conditions" (p. 26)
• The large/small judgment for the upper limit and lower limit values is not per-
formed. An error is not generated if the upper limit and lower limit values are
set in reverse, but the judgment cannot be performed properly.
• Refer to "Judgement results display" (p. 73) regarding comparator judgment
results.
• After settings are complete,
can be used to toggle the judgment
range display to confirm the judgment range.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......