8.9 Message Reference
224
Setting and Query of Threshold for Judging Abnormal Measurement Level
Syntax
Command
Query
:LEVel:CHECk:LIMit
<Numeric Value>
:LEVel:CHECk:LIMit?
<Numeric Value>
= 0.01 to 100.00 (NR2)
Explanation
Command Sets the threshold for detecting abnormal measurement level.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
Query
Returns the threshold for abnormal measurement level as an NR2
numerical value.
Example
Command
Query
Response
:LEVel:CHECk:LIMit 5.00
Sets the threshold for judging abnormal measurement level to 5%.
During a measurement if the amount of change of the monitor values
(voltage monitor value, current monitor value) exceeds ±5% it is
detected as an outside of limit range error.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
:LEVel:CHECk:LIMit?
:LEVEL:CHECK:LIMIT 5.00
(when HEADER ON)
5
(when HEADER OFF)
The threshold for judging abnormal measurement level is set to 5%.
Loading of Specified Panel Number
Syntax
Command
:LOAD
<Numeric Value>
<Numeric Value>
= 1 to 99 (NR1)
Explanation
Command Loads the specified panel number.
A numeric value in NRf format is accepted but decimals are rounded
off so the numeric value can be handled.
Example
Command
:LOAD 2
Loads panel number 2.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......