8.9 Message Reference
186
Setting and Query of Upper Limit and Lower Limit Values of Second
Parameter for BIN Function in Deviation Percent Mode
(Only for Models 3504-50, 3504-60)
Syntax
Command
Query
Response
:BIN:SLIMit:DEViation
<Lower Limit Value>,<Upper Limit Value>
:BIN:SLIMit:DEViation?
<Lower Limit Value>,<Upper Limit Value>
<Lower Limit Value>
= OFF/ Numeric Value from -199999 to 199999 (NR1)
<Upper Limit Value>
= OFF/ Numeric Value from -199999 to 199999 (NR1)
Explanation
Command Sets the upper limit and lower limit values of the second parameter in
deviation percent mode.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
Query
Returns the upper limit and lower limit value settings for the second
parameter in deviation percent mode in order of lower limit value and
upper limit value.
Example
Command
:BIN:SLIMit:DEViation -10,10
Sets -10 for the lower limit value and 10 for the upper limit value of the
second parameter in deviation percent mode.
Query
Response
:BIN:SLIMit:DEViation?
:BIN:SLIMit:DEVIATION -10,10
(when HEADER ON)
-10,10
(when HEADER OFF)
-10 is set for the lower limit value and 10 is set for the upper limit
value of the second parameter in deviation percent mode.
Note
The measurement value for the second parameter in deviation per-
cent mode is the result of the calculation (measurement value - refer-
ence value).
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......