8.9 Message Reference
241
8
Setting and Query of Trigger Mode
Syntax
Command
Query
Response
:TRIGger
<Character>
:TRIGger?
<Character>
<Character>
= INTernal/ EXTernal
INTernal: Internal trigger
EXTernal: External trigger
Explanation
Command Sets the trigger mode.
Query
Returns the setting of the trigger mode as characters.
Example
Command
:TRIGger INTernal
Sets the trigger mode to internal trigger.
Query
Response
:TRIGger?
:TRIGGER INTERNAL
(when HEADER ON)
INTERNAL
(when HEADER OFF)
The trigger mode is set to internal trigger.
Setting and Query of Trigger Delay Time
Syntax
Command
Query
Response
:TRIGger:DELay
<OFF/ Numeric Value>
:TRIGger:DELay?
<OFF/ Numeric Value> = OFF/ 0 to 9.999
(NR2)
Explanation
Command Sets the trigger delay time.
When the trigger delay time is set, the trigger delay function will not be
automatically set to ON.
A numeric value in NRf format is accepted but decimals are rounded
off so the numeric value can be handled.
Query
Returns the trigger delay time as a NR2 numerical value.
Example
Command
:TRIGger:DELay 0.1
Sets so that measurement will commence 100 ms after the trigger.
Query
Response
:TRIGger:DELay?
:TRIGGER:DELAY 0.1
(when HEADER ON)
0.1
(when HEADER OFF)
Measurement is set to commence 100 ms after the trigger.
Note
A command of OFF is accepted, but the setting will become 0, and an
0 response will be returned.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......