5.2 BIN Measurement Function (Model 3504-50, 3504-60 only)
76
Setting Judgment Mode_________________________________________
Set the judgment mode before setting the BIN judgment conditions.
(Select the count value setting, the deviation percent (
%) setting.)
The judgment mode is common to comparator measurement and BIN measure-
ment.
1.
In normal measurement mode, press
.
The upper part of the MAIN display area displays the menu contents and the
lower part displays the setting information.
(Refer to "Menu display organization" (p. 13) for menu order)
This setting cannot be changed in comparator mode and BIN mode.
2.
Use
or
to select the “JudGE” menu item.
3.
Use
or
to select a setting item.
Pressing
or
causes the display to change in the following manner.
4.
Press .
The judgment mode is confirmed.
After confirmation, “bEEP_J”(Beep setting screen for judgment result) is dis-
played at the top of the MAIN display area.
The judgment mode is not confirmed unless
is pressed.
5.
Press .
The unit returns to normal measurement mode.
(Judgment mode setting screen)
Menu content
Judgment mode
(set in step 3)
(MAIN display area)
“Count”
Count setting.
“d-PEr”
Deviation percent (
%) setting.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......