8.2 Specifications
133
8
Control during
Sending
When using hardware (RTS/CTS control):
When CB(CTS) is confirmed to be OFF, the sending of data is halted. When it is
confirmed to be ON, the sending of data is resumed.
When using software (XON/XOFF control):
When XOFF is received, the sending of data is halted. When XON is received,
the sending of data is resumed.
Interface Functions _____________________________________________
Code used: ASCII code
CA(RTS)
OFF
ON
25%
85%
Input Buffer
Usage Amount
Buffer Space
75%
Send XOFF
Send XON
8.2.2 GP-IB Specifications
(Only for Models 3504-50, 3504-60)
SH1
Supports all source handshake functions.
AH1
Supports all acceptor handshake functions.
T6
Supports standard talker functions.
Supports serial poll functions.
Talk only mode is not supported.
Supports the talker cancel function by MLA (My Listen Address).
L4
Supports standard listener functions.
Listener only mode is not supported.
Supports the listener cancel function by MTA (My Talk Address).
SR1
Supports all service request functions.
RL1
Supports all remote/local functions
PP0
Parallel poll functions are not supported.
DC1
Supports all device clear functions.
DT1
Supports all device trigger functions.
C0
Controller functions are not supported.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......