5.2 BIN Measurement Function (Model 3504-50, 3504-60 only)
88
Performing BIN Measurement ____________________________________
• Starting with the lowest, checks to see if BIN numbers are within the set upper
limit and lower limit value range and displays as a judgment result the first BIN
number that is in range.
• Except for the trigger setting, the measurement conditions cannot be changed
in BIN measurement mode. Press
to switch to normal measurement
mode and then change the measurement conditions.
• Press
to toggle the information shown in the SUB display area.
C reference value (for the
% setting)
BIN 1 to 14 upper limit and lower
limit values
D reference value (for the
% setting)
D upper limit and
lower limit values
display OFF
C reference value...
• If the upper limit and lower limit values of C and D have not been set, OUT will
be displayed in the BIN judgment result display area. The measurement range
will automatically become HOLD.
Judgement results
display
Each of the judgment results is displayed in the BIN judgment result display
area.
Hi
Lo
Lo
Hi
OUT
OUT
D-NG
D-NG
C
D
BIN1
BIN2
…
BIN14
When the measurement
value is larger than the
upper limit value of C
When the measurement
value is within the range
of the upper limit and
lower limit values of C
When the measurement
value is smaller than the
lower limit value of C
Upper limit
value
Lower limit
value
When the measurement
value is smaller than the
lower limit value of D
When the measurement
value is larger than the
upper limit value of D
When the measurement value is within
the judgment range of the upper limit
and lower limit values of D
Lower limit value Upper limit value
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......